Cause of Device-Related Incident
*Not stated

Clinical Specialty or Hospital Department
Clinical/Biomedical Engineering; Nursing; Pharmacy / IV therapy

Device Factors
*Not stated

Document Type
Frequently Asked Questions (FAQ)

External Factors
*Not stated

Mechanism of Injury or Death
Embolism (gaseous or particulate)

Support System Failures
*Not stated

Tampering and/or Sabotage
*Not stated

User Errors
*Not stated

UMDNS
Infusion Pumps, General-Purpose [13-215]

Infusion Pump Air Detection



FAQ [Health Devices Apr-May 1998;27(4-5):174-5]

Hospital: What are the appropriate air-detection volumes for infusion pump air-in-line sensors?

ECRI: Air-in-line sensors, which are found in general-purpose infusion pumps, can typically detect minimum air volumes between 0.05 and 0.2 mL. This range is several orders of magnitude smaller than the volume of air thought to be capable of causing a lethal embolus if infused into a patient—common estimates of the lethal volume vary between 10 and 500 mL. Peripheral venous emboli, such as those that are possible during traditional peripheral infusions, usually result from air volumes between 100 and 500 mL. On the other hand, central venous emboli, such as those that can occur during infusion through a peripherally inserted central catheter (PICC), can be produced even by the lower air volumes in the range thought to be lethal.

We are not aware of any reports of fatal venous air embolism that occurred during the use of pumps having functional air-in-line detectors. This suggests that any detector with minimum volumes within the 0.05 to 0.2 mL range provides sufficient protection from an embolism.

Some pumps allow the air-detection volume to be selected from a variety of settings. When these pumps are used, the smallest volume should be selected, especially for neonatal applications. However, this recommendation does not apply to pumps that can be set to detection limits below 0.05 mL, since such limits are likely to cause frequent nuisance alarms. We do not believe that such settings need to be used. If a facility feels a strong need to prevent the infusion of air volumes less than 0.05 mL, an air-eliminating filter can be used downstream from the pump. And some pump administration sets have air traps that will collect typical air volumes produced during infusion, which can further reduce air-in-line risks.

UMDNS Term

Infusion Pumps, General-Purpose [13-215]


[Home]    [About]    [Help]    [Site Map]
Copyright © 2017 ECRI
All rights reserved
www.ecri.org