Hazard [Health Devices Nov 1982;12(1):22-5]
Problem
ECRI received from member hospitals two reports of fatal cerebral air
embolism immediately following routine recalibration of an invasive arterial blood
pressure monitoring system. In both cases, the equipment was calibrated by pressurizing
the system with air using a sphygmomanometer cuff inflation bulb and comparing the
pressure monitor display with the pressure indicated by a mercury manometer. Apparently, a
bolus of air was infused into the patient, either because the stopcock between the
catheter and pressure tubing had not been closed or because air had been inadvertently
introduced into the transducer dome or continuous flushing device, then fast-flushed into
the patient's artery after calibration.
Discussion
Blood pressure transducers and electronic amplifiers have had a history of
calibration problems; therefore, many users have instituted procedures for checking system
calibration to ensure that blood pressures are measured accurately. The monitoring system
is usually calibrated during periodic inspections and immediately before connection to the
patient. This well-known technique compares the electronic pressure display to the
pressure indicated on a mercury manometer (after zeroing the transducer) and is relatively
simple to perform with inexpensive equipment. However, in some hospitals, this same method
is also used to recalibrate the system while it is connected to the patient at eight-hour
intervals (work-shift changes) and after suspicious or clinically significant readings.
Using this technique, the clinician must turn off the stopcock between the transducer and
patient to pressurize the system with the cuff inflation bulb.
Most clinicians recognize that air embolism is possible if this step is
omitted. However, it can be overlooked. When it is, air pressure greater than the blood
pressure will force air into the transducer dome and tubing system and, possibly, into the
patient. Under these conditions, the calibration pressure (typically 100 or 200 mm Hg)
will not be reached, and the clinician may try additional squeezes of the inflation bulb,
further increasing the likelihood of air embolism. Air can also be introduced into the
system or patient if the patient stopcock is opened before the air pressure is released.
Another potential mechanism for air embolus formation is fast-flushing of
the system when air is present, using continuous flush valves. Depending on its duration,
flushing can cause significant retrograde flow in the arterial vessels because the flush
pressures are typically greater than the arterial pressures. Even small amounts of air,
inadvertently introduced into the system, can be flushed from the radial artery, up the
brachial artery and aorta to the cerebral and coronary blood supply, where they can cause
an air embolus. (For example, a small amount of air can be introduced into the system,
even with the patient stopcock closed during calibration, if the calibration pressure
exceeds that of the pressure infuser flushing system. Air might also be inadvertently
introduced after blood sampling.)
In general, the benefits of continuous flush valves to maintain an open
catheter lumen outweigh the risks of retrograde flow during fast-flushing. The likelihood
of significant retrograde flow can be reduced by minimizing the time that the fast-flush
valve is activated. (This is especially relevant when the system's fidelity is checked by
observing the dynamic response to a step function generated by quick-releasing the
fast-flush mechanism.) In any event, it is imperative to keep the system free of air at
all times.
To more safely insure accurate readings while monitoring, most hospitals
rely on periodic rezeroing by opening the dome stopcock to atmosphere and resetting the
zero baseline of the monitor when the transducer is positioned at the level of the heart.
The electronic balance of the monitor, sometimes referred to as electronic
"cal," is also checked. Unfortunately, this function only simulates a transducer
imbalance corresponding to a specified pressure and may not always detect a faulty
transducer. Most monitor manufacturers and some transducer manufacturers recommend
rezeroing and electronic calibration (although one manufacturer's instruction manual still
describes an air-pressurization technique); these procedures have proved to be adequate
for most purposes and present little risk of air embolism. In fact, results of our
evaluation of pressure transducers (Health Devices, Vol. 8, p. 199) support the
adequacy of these procedures since we did not find significant drift in transducer
sensitivities. Transducer failure is most likely to occur during cleaning or other
handling. Sudden transducer failure or significant sensitivity changes are unlikely while
the transducer is undisturbed during monitoring.
We believe that many clinicians have unrealistic expectations for the
accuracy of the monitoring system during arterial blood pressure monitoring. System
accuracy is more limited by the dynamic response of the hydraulic system (tubing,
stopcocks, and catheter) than by the electronic equipment. Resonance in the hydraulic
system can cause overshoot of the pressure pulse, resulting in errors of up to 20 mm Hg in
the systolic pressure value (see Health Devices, Vol. 8, p. 201). Our accuracy
criterion for the monitoring system allows errors of 2 mm Hg or 10%, whichever is less,
when static pressures are applied.
We believe that improvements in pressure transducers and monitors and the
standardization of transducer sensitivities and monitor gain have significantly reduced
the need for routine recalibration of the system during use. Initial zeroing and
calibration and periodic rezeroing and electronic calibration checks are still required.
However, we believe that the risk of air emboli caused by forgetting to close the patient
stopcock during air-pressurized calibration procedures outweighs the risk of inappropriate
diagnosis or therapy from slight system inaccuracies.
Recommendations
- Review hospital procedures for invasive blood
pressure monitoring and criteria for recalibration intervals.
- Eliminate routine recalibration procedures using
air-pressurizing equipment (i.e., cuff inflation bulbs) while the system is
in use.
- If rezeroing and electronic calibration cannot resolve a clinical discrepancy, consider an
alternative recalibration technique (see discussion below) after
discussing system accuracy requirements and infection control policies with
hospital staff.
Alternative In-Use Calibration Techniques
There may be rare instances where a blood pressure monitoring system in use
needs recalibration with a known pressure reference. For example, the
transducer may be jarred or the monitor's gain adjustment (if so equipped) may
be inadvertently changed. A physician may want the system recalibrated before
taking action to confirm that a significant blood pressure change has
occurred.
There are methods of applying reference pressures for those special
circumstances that minimize the risk of air embolism. One technique is to use a
vertically oriented length of tubing connected to the transducer dome and
flushed with solution from the pressure infuser to produce a water manometer.
For example, a saline column within a raised one-meter length of extension
tubing will generate about 73 mm Hg. (Three- and four-foot tubing lengths will
generate pressure of 67 mm Hg and 90 mm Hg, respectively.) This technique can be
performed rapidly and will not introduce air into the system. However, this
calibration pressure may not be adequate for confirmation of high systolic
pressures, and the technique may be awkward in some settings. Slight errors may
result if the tubing is not held vertically.
If disposable diaphragm domes are used, the user can disconnect the transducer
from the sealed hydraulic system and reconnect the transducer to another
dome, which is then used for calibration. After calibration, the transducer
can be reconnected to the hydraulic system, rezeroed, and quickly put back
into operation. However, some manufacturers believe that inconsistent dome couplings
caused by variations in dome application techniques or coupling differences
between dome lots, may result in transducer sensitivity changes of up
to 3%.
Another rarely used technique is to adapt a liquid isolator device for
recalibrating the system. We have had limited success in tests of one such
device. The isolator, a balloon membrane within a syringe barrel, is
connected between the transducer dome and manometer guage system, and the system
is pressurized using the pressure infuser by activating the
"fast-flush" of the continuous flush valve. The manometer and monitor
readings can then be compared at various pressures. This system also
minimizes the risk of air embolism. However, pressure will not be accurately
transmitted to the manometer if the balloon collapses. Also, failure to
close the stopcock to the patient before actiivating the fast-flush mechanism
can cause retrograde flow into the blood vessel or overinfusion of the flush
solution.
Other manufacturers have developed devices for calibration of pressure
monitoring systems. One such device automatically generates air pressures for
calibrating the monitoring system and displays those pressures on a digital
display. this device is appropriate for calibration before use or during
inspection or maintenance, but may cause air embolism if used while the system
is in use and the patient stopcock is not closed.
Another calibration device may be safely used under certain circumstances.
This device uses a piston/cylinder arrangement that generates pressure when the
piston is moved within the cylinder. This mechanism limits the total volume of
air that can be introduced. However, the risk of air embolism does exist if
this volume is greater than the volume within certain monitoring kits (i.e.,
when short tubing lengths are used with miniature transducers mounted on the
patient's limb). The risk is much less for IV pole-mounted systems with longer
tubing lengths, which have greater volumes. Users of this device should also
consider that it generate expected pressures based on compression of a
"known" volume of air and cannot be used as a second pressure standard
(i.e., like a mercury manometer) to compare with the monitor reading. Unexpected
compliances (e.g., caused by excessive microbubbles or air-filled tubing lengths
used to connect the device and system) may cause calibration errors with this
device.
The recalibration technique must also minimze the risk of contaminating the
systems's sterile fluid pathway. Stopcock and dome ports can become contaminated
using any recalibration or rezeroing technique. It is usually desirable to
interface calibration equipment and the monitoring system with a low-cost
sterile component that can be discarded later, such as a stopcock, short length
of tubing, or filter designed for this purpose. However, the use of a filter
will not necessarily reduce the risk of air embolism when air pressurization
calibration is performed. The flush solution manometer (e.g., PVC tubing) and
the fluid isolator device use prepackaged sterilized disposables, and,
therefore, aid in reducing the risk of contamination during
recalibration.
References
Edmonds JF, Barker GA, Conn AW. Current concepts in cardiovascular
monitoring in children. Crit Care Med 1980;8(Oct):548-553.
Lantiegne KC, Civetta JM. A system for maintaining invasive pressure
monitoring. Heart Lung
1978;7(Jul-Aug):610-21.
Lowenstein E, Little JW, Lo HH. Prevention of cerebral embolization from
flushing radial-arterial cannulas. N Engl J Med 1971;285(Sep):1414.
Smith RN. Invasive pressure monitoring. Am J Nurs
1978;78(Sep):1514-21.
UMDNS Term
Pressure Monitors, Blood, General/Invasive [16-764]
Cause of Device-Related Incident
Device factor: Device interaction
User errors: Inappropriate reliance on an automated feature; Incorrect clinical use
Support system failure: Lack or failure of incoming and pre-use inspections
Mechanism of Injury or Death
Embolism (gaseous)