Disposable Blood Pressure Transducers: Calibration Methods
User Experience Network™ [Health Devices Feb 1993;22(2):97]
Hospital
We are currently looking at different suppliers from whom to obtain
disposable blood pressure transducers. Several models that we are looking at do not have
the back-side calibration feature that our current models have. Can you provide us with
your opinion on those models that do not have the back-side calibration port?
ECRI
In our previous evaluation of disposable blood pressure transducers (Health
Devices 17[3], Mar 1988), we slightly preferred units that had a back-side calibration
port over units that lacked this feature. The back-side calibration port simplifies both
pre-use testing of the transducer and accuracy verification of transducers that are
already connected to a patient. Units that have such a port often require a special
pneumatic test device. Because this device costs about $500, hospitals should negotiate
with vendors to minimize any additional cost when contracting for new transducers. Users
should also check the pneumatic tester's accuracy at least twice a year.
Do not use a mercury or aneroid manometer to test the transducer after it
is connected to the patient. Errors such as incorrect stopcock positioning can result in
pumping air into the blood vessel. We have received reports of fatal cerebral air embolism
immediately following routine recalibration of invasive arterial blood pressure monitoring
systems while connected to a patient (see our Hazard Report in Health Devices
12[1], Nov 1982).
Problems with pressure monitoring accuracy are likely to be caused by
blood clotting at the tip of the catheter or poor dynamic response of the pressure
monitoring system. Questions also arise when noninvasive (cuff) and invasive (transducer)
measurements do not agree; however, because of differences in the measurement method and
location of the measurement within the circulatory system, these results will often
differ.
Pre-use testing of transducers will virtually eliminate the possibility
that a transducer failure will cause an erroneous measurement. In the few instances in
which a concern exists about a transducer that is already in use, the transducer can
simply be replaced (after checking zeroing) if it does not have a back-side calibration
port. This would further minimize risks and eliminate the time expenditure needed to
perform such calibration. If the unit has a back-side calibration port, the calibration
procedure will be easier and more reliable.
UMDNS Term
Transducers, Pressure, Disposable [16-909]
Cause of Device-Related
Incident
Device factor: Improper maintenance, testing,
repair, or lack or failure of incoming inspection
Mechanism of Injury or
Death
Embolism (gaseous); Monitoring failure