Scavenging Gas from Membrane Oxygenators
Hazard [Health Devices Sep-Oct 1987;16(9-10):343-4]
Problem
While investigating a heart-lung bypass accident involving gas embolism,
ECRI discovered that some perfusionists use what may be unsafe techniques to scavenge
waste gas from membrane oxygenators. In this accident, it appeared that the technique used
to scavenge waste gas may have been responsible for generating back pressure in the gas
jacket surrounding a microporous hollow-fiber membrane oxygenator. This back pressure can
force gas through the fibers and into the blood flowing through them. The magnitude of the
back pressure dictates the volume of gas introduced into the blood and may determine the
extent of the sequelae. In this case, the patient died as a result of the embolism.
Discussion
Although membrane oxygenators have been in existence since heart-lung
bypass was introduced, the development of efficient gas-exchange membranes and the
realization that direct contact with gas may traumatize blood have led to increased use of
membrane oxygenators. Unlike bubble oxygenators, which rely on direct contact of blood and
bubbles for gas exchange and are designed to separate undissolved gas from blood before
the blood exits the oxygenator, membrane oxygenators are not designed to separate blood
and bubbles of undissolved gas. Consequently, large volumes of undissolved gas entering
with blood or being generated by back pressure in the membrane oxygenator will flow out of
the oxygenator with the blood.
At the reporting hospital, a ¼-inch ID section of tubing had been
connected to the gas outlet port of the hollow-fiber membrane oxygenator to scavenge waste
anesthetic gas from the oxygenator. This section of tubing was in turn connected to a
vacuum line. Several ½-inch diameter holes were cut in the tubing at evenly spaced
intervals to prevent the oxygenator from being exposed to excessive negative pressure from
the vacuum, which could possibly result in poor blood oxygenation. We believe that at some
point during the bypass procedure, the gas scavenging tubing became kinked at the location
of the hole nearest to the oxygenator gas vent port, causing back pressure to develop in
the gas jacket of the oxygenator. Ironically, the holes cut in the tubing to prevent one
problem may have caused another. As a result of the holes cut in the tubing wall, the
tubing in the vicinity of the holes was less resistant to kinking and occlusion than
sections of tubing where the wall was intact.
Although this accident may have been the result of a unique set of
circumstances, it raises the concern that other hospitals may not be scavenging anesthetic
gases safely from membrane oxygenators. In tests conducted in our laboratory on the same
type of oxygenator that was used by the reporting hospital, we found that oxygen at a
pressure of 26 mm Hg in the gas jacket could produce bubbles in the blood-carrying
channels of the hollow-fiber oxygenator. Oxygen supplied from tanks or piped into the OR
is typically delivered at a regulated pressure of 50 psi (2,600 mm Hg), and hospitals
routinely connect gas lines from these sources to membrane oxygenators through gas
blenders or flowmeters that do not introduce significant pressure drops in the gas.
Clearly, the potential for gas embolism exists if the gas vent port of these oxygenators
becomes either partially or totally occluded. Gas scavenging systems for these oxygenators
must not cause the application of positive or negative pressures in the gas jackets as
this may adversely affect the patient.
The American National Standards Institute (ANSI) standard (ANSI Z79.11) addressing scavenging systems
for anesthetic gases applies to oxygenators as well as to anesthesia machines. This
standard states that scavengers must not generate positive pressures exceeding 10 cm of
water (7.4 mm Hg) or negative pressures exceeding 0.5 cm of water (0.37 mm Hg). While some
manufacturers have informed ECRI that the design of their membrane oxygenators will
prevent the development of back pressure in the gas jacket, we believe that precautions
should still be taken to ensure that the gas scavenging system cannot adversely affect any
membrane oxygenators.
Recommendations
- Be sure to consult manufacturer's instructions when
setting up a scavenging system for membrane oxygenators. Contact the
manufacturer to ensure that the recommended scavenging technique meets the
ANSI performance standard.
- Do not exceed the gas flow rates recommended by the
manufacturer for a particular oxygenator as excessive flow rates may cause
excessive back pressure in the gas jacket of oxygenators.
- If the manufacturer does not provide instructions on how to scavenge
waste anesthetic gas from membrane oxygenators, use an anesthesia machine scavenging
system that provides active scavenging (non-HVAC type) and meets the ANSI performance
standard. (For more information on anesthetic gas scavengers, see Health Devices 12:267-80, September 1983.)
- The section of tubing between the oxygenator and
the scavenger must also prevent positive pressure from developing in the
oxygenator. Be sure that this tubing is as short as possible to prevent
kinking. When installing scavenging systems near oxygenators, verify that
the scavenger's orientation will not affect its performance. If the
manufacturer recommends the installation of a rigid plastic T-tubing
connector or an open Luer-lock fitting in the tubing between the oxygenator
and the gas scavenging system to provide positive pressure relief, be sure
to install it as close to the oxygenator as possible.
- Route all tubing, including blood lines, for a heart-lung bypass system
in such a way that it will not be easily kinked, walked on, or pinched as
equipment is moved near the pump console. Never cover or drape tubing so
that it cannot be monitored by the perfusionist during bypass.
UMDNS Terms
- Anesthesia Unit Gas Scavengers [10-142]
- Oxygenators, Extracorporeal, Membrane [17-643]
Cause of Device-Related Incident
Device factors: Design/labeling error;
Device interaction; Improper maintenance
User errors: Failure to read label; Incorrect clinical use
External factor: Medical gas and vacuum supplies
Mechanism of Injury or Death
Embolism (gaseous)